The reliability and potential value of a specific ‘centre of pressure locator’ in chiropractic practice

This study assessed the reliability and potential value of a specific Centre of Pressure Locator (COPL) for the initial diagnosis of spinal mal-alignments and for the measurement of change in weight distribution resulting from clinical intervention. Basic validation of the equipment with standard weights showed it to be very precise, reliable and accurate at noting changes in the position of the centre of pressure. Control subjects were used to develop interim norms for COP position and sway. R-L COP position among both controls and patients was found to be too variable to be a useful tool for diagnosis or for the measurement of the effects of intervention. However, the equipment shows promise for the measurement of A-P and R-L postural sway; potentially important variables to consider within chiropractic practice. (JCCA 2000; 44(4):209–222)
Key Words: chiropractic, rehabilitation, exercise

La présente étude vise à évaluer la fiabilité et la valeur potentielle d’un localisateur du centre de gravité (CG) pour le diagnostic initial des défauts d’alignement de la colonne vertébrale et la mesure des changements dans la répartition du poids suite aux interventions cliniques. Des tests élémentaires de validation, réalisés à l’aide de poids standards ont révélé que le matériel est précis, fiable et exact pour ce qui est des changements de position du centre de gravité. Des sujets témoins ont servi à l’élaboration de normes provisoires sur la position et le déplacement du CG. Les écarts chez les patients pour que le matériel s’avère utile à la pose d’un diagnostic ou à la mesure des effets des interventions. Par contre, il se montre prometteur pour la mesure des déplacements posturaux antéro-postérieurs ou droite-gauche, deux variables. JACC 2000; 44(4):209–222)
Mots-clés: chiropratique, réadaptation, exercices

back to top